Feasibility and accuracy of relative electron density determined by virtual monochromatic CT value subtraction at two
Toshiyuki Ogata1, Takashi Ueguchi, Masashi Yagi
1Department of Radiology, Osaka University Hospital, 2-15 Yamadaoka, Suita, Osaka, Japan. ogata@k-mcc.net
Background:
Recent work by Saito (2012) has demonstrated a simple conversion from energy-subtracted computed tomography (CT) values (ΔHU) obtained using dual-energy CT to relative electron density (RED) via a single linear relationship. The purpose of this study was to investigate the feasibility of this method to obtain RED from virtual monochromatic CT images obtained by the gemstone spectral imaging (GSI) mode with fast-kVp switching.
Methods:
A tissue characterization phantom with 13 inserts made of different materials was scanned using the GSI mode on a Discovery CT750 HD. Four sets of virtual monochromatic CT images (60, 77, 100 and 140 keV) were obtained from a single GSI acquisition. When we define Δ HU in terms of the weighting factor for the subtraction α, Δ HU ≡ (1 + α)H - αL (H and L represent the CT values for high and low energy respectively), the relationship between Δ HU and RED is approximated as a linear function, a × Δ HU/1000 + b (a, b = unity). We evaluated the agreement between the determined and nominal RED. We also have investigated reproducibility over short and long time periods.
Results:
For the 13 insert materials, the RED determined by monochromatic CT images agreed with the nominal values within 1.1% and the coefficient of determination for this calculation formula was greater than 0.999. The observed reproducibility (1 standard deviation) of calculation error was within 0.5% for all materials.
Conclusions:
These findings indicate that virtual monochromatic CT scans at two different energies using GSI mode can provide an accurate method for estimating RED.
More Related Videos
07:24Quantitative Atomic-Site Analysis of Functional Dopants/Point Defects in Crystalline Materials by Electron-Channeling-Enhanced Microanalysis
Published on: May 10, 2021
10:42In Depth Analyses of LEDs by a Combination of X-ray Computed Tomography (CT) and Light Microscopy (LM) Correlated with Scanning Electron Microscopy (SEM)
Published on: June 16, 2016
Related Concept Videos
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...