Oscillometric estimation of aortic pulse wave velocity: comparison with intra-aortic catheter measurements
Bernhard Hametner1, Siegfried Wassertheurer, Johannes Kropf
1Department of Health and Environment, AIT Austrian Institute of Technology, Vienna, Austria.
Objectives:
Recently, a novel method to estimate aortic pulse wave velocity (aPWV) noninvasively from an oscillometric single brachial cuff waveform reading has been introduced. We investigated whether this new approach provides acceptable estimates of aPWV compared with intra-aortic catheter measurements.
Methods:
Estimated values of aPWV obtained from brachial cuff readings were compared with those obtained using an intra-aortic catheter in 120 patients (mean age 61.8±10.8 years) suspected for coronary artery disease undergoing cardiac catheterization. Differences between aPWV values obtained from the test device and those obtained from catheter measurements were estimated using Bland-Altman analysis.
Results:
The mean difference±SD between brachial cuff-derived values and intra-aortic values was 0.43±1.24 m/s. Comparison of aPWV measured by the two methods showed a significant linear correlation (Pearson's R=0.81, P<0.0001). The mean difference for repeated oscillometric measurements of aPWV was 0.05 m/s, with 95% confidence interval limits from -0.47 to 0.57 m/s.
Conclusion:
aPWV can be obtained using an oscillometric device with brachial cuffs with acceptable accuracy compared with intra-aortic readings.
More Related Videos
05:51Measuring the Carotid to Femoral Pulse Wave Velocity (Cf-PWV) to Evaluate Arterial Stiffness
Published on: May 3, 2018
06:04Pulse-Wave Velocity, Flow-Mediated Dilation, and Carotid Intima-Media Thickness to Assess Cardiovascular Risk in Population with Metabolic Syndrome
Published on: September 27, 2024
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pulse
The pulse serves as a clinical indicator...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...