Forces generated by Macintosh and GlideScope® laryngoscopes in four airway-training manikins
1University of Toronto, Toronto General Hospital, Toronto, ON, Canada. corinalee99@gmail.com
Abstract:
We measured forces generated by Macintosh and GlideScope(®) laryngoscope blades during airway intubation by 16 participants in four manikins: Laerdal(®) SimMan; TruCorp AirSim™ Advance; Laerdal(®) Airway Management Trainer; and Ambu(®) Airway Man. Both laryngoscopes generated the least force in the Laerdal Airway Management Trainer and the most in the Ambu Airway Man. The respective median (IQR [range]) forces generated by the Macintosh blade were 2 (1-4 [1-7]) N vs 9 (7-13 [5-16]) N, p = 0.00004, with peak forces 9 (5-11 [3-16]) N vs 18 (12-22 [3-31]) N, p = 0.0004. The respective average and peak forces generated by the GlideScope blade were 1 (1-2 [0-3]) N vs 4 (3-5 [2-6]) N, p = 0.00001, and 4 (2-7 [0-12]) N vs 7 (4-9 [3-18]) N, p = 0.054.
Related Concept Videos
Electromotive Force
Intermolecular Forces
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Intermolecular vs Intramolecular Forces
Intermolecular Forces in Solutions
When the strengths of the intermolecular forces of attraction between solute and solvent species in a solution are no different than those present in the separated components, the solution is formed with no accompanying energy change. Such a solution is called an ideal solution. A mixture of ideal gases (or gases such as helium and argon,...