Related Experiment Video
Updated: May 12, 2026

Electrolytic Inferior Vena Cava Model (EIM) of Venous Thrombosis
Published on: July 12, 2011
Lead thrombi associated with externalized cables on Riata ICD leads: a case series
Sandeep K Goyal1, Christopher R Ellis, Jeffery N Rottman
1Vanderbilt Heart and Vascular Institute, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
Abstract:
St. Jude Riata family ICD leads are subject to an FDA class I recall due to insulation failure, cable extrusions, and a high rate of electrical failures. We present multiple cases of large intravascular thrombus formation adherent to externalized conductor cables on Riata leads. Our observations highlight a previously unknown risk of large or calcified thrombus formation involving externalized conductors. We suggest that these leads should be screened for thrombi prior to potential lead extraction to determine the safest approach. Serial echocardiography may help detect progressive thrombus formation. The role of systemic anticoagulation for prevention or treatment of these thrombi remains unclear.
Related Concept Videos
Venous Thrombosis III: Interprofessional Care
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care
Venous Thrombosis I: Introduction