Evaluation of a storage cabinet for heat-sensitive endoscopes in a clinical setting
P Grandval1, G Hautefeuille, B Marchetti
1Endoscopy Department, La Timone University Hospital, Marseille Cedex 5, France.
Background:
In most countries, endoscopes must be disinfected or fully reprocessed before the beginning of each session, even if they were cleaned and disinfected after their last use. Several storage cabinets for heat-sensitive endoscopes (SCHE) are commercially available. They are designed to maintain the microbiological quality of reprocessed endoscopes for a predefined period of time validated by the SCHE manufacturer. Use of an SCHE increases the acceptable storage time before it is necessary to re-disinfect the endoscope.
Aim:
To evaluate the efficacy of an SCHE (DSC8000, Soluscope, SAS Marseilles, France) in a clinical setting.
Method:
The microbiological quality of endoscopes was assessed after 72 h of storage in an SCHE (Group I), and compared with the microbiological quality of endoscopes stored for 72 h in a clean, dry, dedicated cupboard without morning disinfection (Group II) and the microbiological quality of endoscopes stored for 72 h in a clean, dry, dedicated cupboard with morning disinfection (Group III). Forty-one endoscopes in each group were sampled for microbiological quality. Endoscope contamination levels were analysed according to guidelines published by the National Technical Committee on Nosocomial Infection in 2007.
Findings/Conclusion:
Use of an SCHE helps to maintain the microbiological quality of endoscopes, provided that staff members are well trained and all practices are framed by a proven quality assurance process.
Related Concept Videos
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Sample Handling
Samples should be transported carefully from collection points to the laboratory. They should be properly sealed and clearly labeled to prevent cross-contamination. To preserve the sample integrity, optimal temperature conditions during transport are essential. This could involve using...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...