Self-test web-based pure-tone audiometry: validity evaluation and measurement error analysis.
Marcin Masalski1, Tomasz Kręcicki
1Department and Clinic of Otolaryngology, Head and Neck Surgery, Wroclaw Medical University, Wrocław, Poland. marcin.masalski@pwr.wroc.pl
Web-based pure-tone audiometry shows potential for hearing screening tests. Measurement errors were analyzed, with factors like calibration and procedure significantly influencing results, suggesting future improvements for broader application.
Area of Science:
- Audiology
- Digital Health
- Medical Measurement
Background:
- Self-administered Web-based pure-tone audiometry (WBPTA) offers a potential home-based hearing test method.
- The feasibility of WBPTA depends critically on understanding and minimizing measurement error.
- Accurate hearing threshold determination is essential for effective audiological assessment.
Purpose of the Study:
- To determine the measurement error associated with home-based WBPTA.
- To identify and analyze the specific factors contributing to this measurement error.
- To evaluate the reliability of WBPTA compared to clinical audiometry.
Main Methods:
- Hearing threshold evaluation involved three series: clinical audiometry, supervised computer-based self-tests, and unsupervised home-based WBPTA.
- The study included 51 participants from an audiology outpatient clinic.
- 37 participants (73%) completed the home-based self-administered tests.
Main Results:
- High correlation (r=0.90) observed between clinical and supervised self-tests with a small average difference (-1.54dB ± 7.88dB).
- Correlation between clinical and home-based WBPTA was also strong (r=0.84) with a similar average difference (-1.35dB ± 10.66dB).
- Key error sources in home-based WBPTA included hearing threshold identification procedure (6.64dB), calibration (6.19dB), and frequency nonlinearity at 250Hz (7.28dB).
Conclusions:
- Web-based pure-tone audiometry is suitable for hearing screening applications.
- Minimizing measurement errors through future method modifications can expand WBPTA's utility.
- Further research can optimize WBPTA for wider clinical and remote audiological use.