The performance determination of a Compton-suppression spectrometer and the measurement of the low level radioactive
Yuan-Qing Fan1, Shi-Lian Wang, Qi Li
1CTBT Beijing National Data Centre and Radionuclide Laboratory, China. yuanqing.fan@nrl.org.cn
Abstract:
The performance of a new Compton-suppression spectrometer consisting of one HPGe detector and three NaI(Tl) detectors was studied. The peak-to-Compton ratio for a (137)Cs source is 1150 and the integral background count rate is 0.3 5s(-1) over the energy interval 20-3000 keV. The spectrometer was used to acquire both Compton-suppressed and non-suppressed spectra of aerosol samples collected in Beijing following the Fukushima nuclear accident.