Sinogram affirmed iterative reconstruction in head CT: improvement of objective and subjective image quality with
1Department of Diagnostic und Interventional Neuroradiology, Hoppe-Seyler-Straße 3, 72076 Tübingen, Germany.
Purpose:
Iterative reconstruction has recently been revisited as a promising concept for substantial CT dose reduction. The purpose of this study was to assess the potential benefit of sinogram affirmed iterative reconstruction (SAFIRE) in head CT by comparing objective and subjective image quality at reduced tube current with standard dose filtered back projection (FBP).
Materials And Methods:
Non-contrast reduced dose head CT (255 mAs, CTDIvol 47.8 mGy) was performed in thirty consecutive patients and reconstructed with SAFIRE and FBP. Images were assessed in terms of quantitative and qualitative image quality and compared with FBP of standard dose acquisitions (320 mAs, CTDI vol 59.7 mGy).
Results:
In reduced dose CT examinations, use of SAFIRE versus FBP resulted in 47% increase in contrast-to-noise ratio (CNR) (2.49 vs. 1.69; p<0.0001). While reduction of tube current was associated with 13% decrease in CNR, quantitative degradation of image quality at lower dose was more than compensated through SAFIRE (2.49 vs. 1.96; p=0.0004). Objective measurements of image sharpness were comparable between FBP and SAFIRE reconstructions (575.9 ± 74.1 vs. 583.4 ± 74.7 change in HU/Pixel; p=0.28). Compared to standard dose FBP, subjective grading of noise as well as overall image quality scores were significantly improved when SAFIRE was used in reduced dose exams (1.3 vs. 1.6, p=0.006; 1.3 vs. 1.7, p=0.026).
Conclusion:
At 20% dose reduction, reconstruction of head CT by SAFIRE provides above standard objective and subjective image quality, suggesting potential for more vigorous dose savings in neuroradiology CT applications.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT