Accuracy of distal radius positioning using an anatomical plate
Related Experiment Video
Updated: May 12, 2026

Pedicle Screw Placement Using an Augmented Reality Head-Mounted Display in a Porcine Model
Published on: May 24, 2024
Joy C Vroemen1, Johannes G G Dobbe, Inger N Sierevelt
1Department of Plastic, Reconstructive and Hand Surgery, Academic Medical Center, University of Amsterdam, The Netherlands. joyvroemen@hotmail.com
Abstract:
Over the past decade, several anatomical plates have been introduced to improve the result of open reduction and internal fixation of the distal radius. Using 3-dimensional imaging techniques, the authors studied the accuracy and reproducibility of distal radius positioning using anatomical plates.Distal radius fractures and the correction of these fractures were simulated with plastic bone models of radii. The authors simulated a defect by removing an arbitrary wedge shape from the artificial radii. Two surgeons corrected these fractures by placing 2 anatomical plate types according to the plate manufacturers' instructions. The residual positioning errors of the distal segment in relation to the unaffected radii were determined using 3-dimensional imaging and were compared with naturally occurring bilateral radius differences in healthy individuals. In many cases, positioning does not agree with differences based on bilateral asymmetry in healthy patients.This study indicated the accuracy of anatomical plates. Positioning an anatomical plate may lead to considerable residual errors in individual patients. Volar distal radius plate shapes differ among plate manufacturers. Therefore, one plate may perform better than another in an individual.