Evaluation of vancomycin susceptibility testing for methicillin-resistant Staphylococcus aureus: comparison of Etest
Michael J Rybak1, Celine Vidaillac, Helio S Sader
1Anti-Infective Research Laboratory, Eugene Applebaum College of Pharmacy & Health Sciences, Detroit, Michigan, USA. m.rybak@wayne.edu
Abstract:
We evaluated the ability of four commercial MIC testing systems (MicroScan, Vitek 2, Phoenix, and Etest) to detect vancomycin MIC values of ≤1 to ≥2 in 200 methicillin-resistant Staphylococcus aureus (MRSA) strains compared to the Clinical and Laboratory Standards Institute broth microdilution (BMD) reference methods. Compared to the BMD method, absolute agreement (0 ± dilution) was highest for the Phoenix system (66.2%) and the MicroScan turbidity method (61.8%), followed by the Vitek 2 system (54.3%). The Etest produced MIC values 1 to 2 dilutions higher than those produced by the BMD method (36.7% agreement). Of interest, the MicroScan system (prompt method) was more likely to overcall an MIC value of 1 mg/liter (74.1%), whereas the Phoenix (76%) and Vitek 2 (20%) systems had a tendency to undercall an MIC of 2 mg/liter. The ability to correctly identify vancomycin MIC values of 1 and 2 has clinical implications and requires further evaluation.
Related Concept Videos
Automated Microbial Diagnostics
Antimicrobial Effectiveness
Clinical Significance of Antibiotic Resistance
Mechanism of Antibiotic Resistance in MRSA
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...