Background And Objective:
Nocturnal application of continuous positive airway pressure (CPAP) is the standard treatment for patients with obstructive sleep apnoea (OSA). Determination of the therapeutic pressure (CPAP titration) is usually performed by a technician in the sleep laboratory during attended polysomnography. One possible alternative to manual titration is automated titration. Indeed, during the last 15 years, devices have been developed that deliver autoadjustable CPAP (A-CPAP). The aim of the present study was to compare the titration effectiveness of two A-CPAP devices using different flow-based algorithms in patients with OSA.
Methods:
This is a randomized study; 79 subjects underwent two consecutive unattended home A-CPAP titration nights with two different devices (Autoset Resmed; Remstar Auto Respironics); during the third and the fourth night, patients underwent portable monitoring in the sleep laboratory during fixed CPAP at the A-CPAP recommended pressure.
Results:
Bland Altman plots showed good agreement between the recommended median and maximal pressure levels obtained with the two devices. A significant improvement was observed in all the sleep parameters by both A-CPAP machines to a similar degree.
Conclusions:
It was observed that the two A-CPAP devices using different algorithms are equally effective in initial titration of CPAP.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Sleep Apnea
The condition is more prevalent among...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Cardiopulmonary Resuscitation II: ACLS Airway Management