Related Experiment Video
Updated: May 12, 2026

Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
Number of basis images effect on detection of root fractures in endodontically treated teeth using a cone beam
Boulos Bechara1, C Alex McMahan, Ibrahim Nasseh
1Oral and Maxillofacial Radiology, Department of Comprehensive Dentistry, University of Texas Health Science Center at San Antonio, San Antonio, TX, USA. boulosbchara@hotmail.com
Objectives:
The objective of this study was to compare the accuracy of root fracture detection in endodontically treated teeth using scans acquired with a 180° rotation motion compared with a 360° after which the patient exposure and number of basis images are doubled.
Methods:
Sixty-six roots were collected and decoronated. All were treated endodontically. One-half of the roots were fractured, resulting in 2 root fragments which were then glued together. The roots were placed randomly in 8 prepared beef rib fragments. Five reviewers independently reviewed the scans twice, at different times.
Results:
The specificity of the 360° scan was significantly higher than the 180° scan; doubling the basis images leads to a significant decrease in false-positive rates. Accuracy and sensitivity were not significantly different.
Conclusions:
Only the specificity is improved by the increased rotation and doubling of images. The accuracy and sensitivity are not improved.