[Analytical method and comparison for static and dynamic headspace gas chromatography of anisole in water]
Yan Zhang1, Jie-feng Qian, Lan-xia Liu
1Physical and Chemical Laboratory, Xuhui, Center for Disease and Prevention, Shanghai 200237, China.
Objective:
To establish and compare the method of static headspace gas chromatography hydrogen flame detector (static headspace method) and purge and trap gas chromatography-mass spectrometry (dynamic headspace method) of anisole in water.
Methods:
Nitrogen gas was used as carrier gas in the static headspace method, 5 g NaCl as matrix modifier was added into 10 ml water. The sample was balanced with high speed vibration at 75°C for 30 min, and anisole was detected by gas chromatography and quantified with external standard. Helium was used as carrier gas in dynamic headspace method, 5.0 ml water and 0.004 mg/L internal standard fluorobenzene was purged into the purge and trap apparatus. After purging, trapping and desorption, anisole was detected by the gas chromatography-mass spectrograph, confirmed by the retention time and comparison of mass-spectrogram in spectrum library and quantified with internal standard. The repeatability and sensitivity of assay were evaluated.
Results:
A good linear range for anisole was observed in static headspace gas chromatography and dynamic headspace gas chromatography-mass spectrometry, within the range of 10 - 500 µg/L and 0.5 - 60.0 µg/L respectively. The linear regression equation was Y = 782.150X + 1.3446 and Y = 0.0358X - 0.0209 respectively, both the correlation coefficient ≥ 0.999. The detection limit (LOD) were 0.002 µg/L and 0.110 µg/L, the lower limit of quantitation (LOQ) were 0.006 µg/L and 0.350 µg/L, the relative standard deviation (RSD) were 1.8% - 2.3% and 2.0% - 3.4%, and the spiking recovery were 93% - 101% and 96% - 101% respectively.
Conclusion:
The methods of static headspace gas chromatography and dynamic headspace gas chromatography-mass spectrometry are simple and can measure anisole in water quickly, sensitively and accurately.
More Related Videos
04:19An Improved Technique for Trimethylamine Detection in Animal-Derived Medicine by Headspace Gas Chromatography-Tandem Quadrupole Mass Spectrometry
Published on: March 10, 2023
10:34Characterization and Application of Passive Samplers for Monitoring of Pesticides in Water
Published on: August 3, 2016
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Chromatographic Methods: Terminology