Calibration of the KRISS reference ionization chamber for certification of ²²²Rn gaseous sources
1Korea Research Institute of Standards and Science, Daejeon 305-340, Republic of Korea. jmlee@kriss.re.kr
Abstract:
A primary measurement system for gaseous (222)Rn based on the defined solid angle counting method has recently been constructed at KRISS and the reference ionization chamber used to measure the activities of gamma-emitting single radionuclides was adopted as a secondary standard for gaseous (222)Rn. A 20 mL flame-sealed glass ampoule source from the primary measurement system was used to calibrate the ionization chamber for (222)Rn. The (222)Rn efficiency of the ionization chamber was compared with that calculated by using a photon energy-dependent efficiency curve and that measured by using a standard (226)Ra solution. From the comparisons we draw the conclusion that the reference ionization chamber for gamma-emitting radionuclides can be a suitable secondary measurement system for gaseous (222)Rn sources.
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...