Accuracy of iodine removal using dual-energy CT with or without a tin filter: an experimental phantom study
Tatsuya Kawai1, Mitsuru Takeuchi, Masaki Hara
1Department of Radiology, Nagoya City University Graduate School of Medical Sciences, Nagoya.
Background:
The effects of a tin filter on virtual non-enhanced (VNE) images created by dual-energy CT have not been well evaluated.
Purpose:
To compare the accuracy of VNE images between those with and without a tin filter.
Material And Methods:
Two different types of columnar phantoms made of agarose gel were evaluated. Phantom A contained various concentrations of iodine (4.5-1590 HU at 120 kVp). Phantom B consisted of a central component (0, 10, 25, and 40 mgI/cm(3)) and a surrounding component (0, 50, 100, and 200 mgI/cm(3)) with variable iodine concentration. They were scanned by dual-source CT in conventional single-energy mode and dual-energy mode with and without a tin filter. CT values on each gel at the corresponding points were measured and the accuracy of iodine removal was evaluated.
Results:
On VNE images, the CT number of the gel of Phantom A fell within the range between -15 and +15 HU under 626 and 881 HU at single-energy 120 kVp with and without a tin filter, respectively. With attenuation over these thresholds, iodine concentration of gels was underestimated with the tin filter but overestimated without it. For Phantom B, the mean CT numbers on VNE images in the central gel component surrounded by the gel with iodine concentrations of 0, 50, 100, and 200 mgI/cm(3) were in the range of -19-+6 HU and 21-100 HU with and without the tin filter, respectively.
Conclusion:
Both with and without a tin filter, iodine removal was accurate under a threshold of iodine concentration. Although a surrounding structure with higher attenuation decreased the accuracy, a tin filter improved the margin of error.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT