Related Experiment Video
Updated: May 11, 2026

10:10
Reverse Total Shoulder Arthroplasty
Published on: July 5, 2011
Fully uncemented glenoid component in total shoulder arthroplasty.
Lieven De Wilde1, Nader Dayerizadeh, Francis De Neve
1Department of Orthopaedic Surgery and Traumatology, Ghent University Hospital, Gent, Belgium.
Journal of Shoulder and Elbow Surgery
|April 27, 2013
Summary
Cementing an all-polyethylene glenoid component in total shoulder arthroplasty may not be necessary. This study found promising clinical and radiologic results with uncemented glenoid components, with 88% showing no loosening on CT scans.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Radiology
Background:
- Glenoid component loosening is a primary complication in total shoulder arthroplasty.
- All-polyethylene components with cemented peg fixation have shown better biomechanical and clinical performance.
- Potential side effects associated with cementing necessitate exploring alternative fixation methods.
Purpose of the Study:
- To evaluate the necessity of cementing a specific all-polyethylene glenoid component with 3 peripheral pegs and 1 central anchor peg.
- To determine if uncemented fixation can achieve comparable clinical and radiologic outcomes.
- To assess the long-term viability of uncemented glenoid components in shoulder arthroplasty.
Main Methods:
- A cohort of 34 patients (34 shoulders) underwent total shoulder arthroplasty with an uncemented all-polyethylene glenoid component.
- Clinical outcomes were assessed using the Constant-Murley score and the SF-12 Health Survey.
- Glenoid component fixation was evaluated using computed tomography (CT) scans at a mean follow-up of 28.3 months.
Main Results:
- The Constant-Murley score significantly improved postoperatively (72 points) compared to preoperatively (40.2 points).
- SF-12 scores indicated satisfactory physical (45) and mental (50) health outcomes.
- CT scans revealed no signs of loosening in 88% (30/34) of the shoulders; loosening was observed around pegs in 4 shoulders, with no significant clinical outcome difference.
Conclusions:
- Uncemented all-polyethylene glenoid components demonstrate promising clinical and radiologic results in total shoulder arthroplasty.
- The majority of patients (88%) showed no signs of loosening on CT scans, suggesting successful uncemented fixation.
- This approach may offer a viable alternative to cemented fixation, potentially mitigating cementing-related complications.
Keywords:
Case SeriesLevel IVTotal shoulder arthroplastyTreatment Studyglenoid componentpolyethyleneuncemented