Related Experiment Video
Updated: May 11, 2026

Correction of Presbyopia by Monocular Bi-Aspheric Ablation Profile
Published on: September 20, 2024
Essential opaque bubble layer elimination with novel LASIK flap settings in the FS200 Femtosecond Laser
A John Kanellopoulos1, George Asimellis
1Laservision.gr Eye Institute, Athens, Greece ; New York University School of Medicine, New York, NY, USA.
Background:
The purpose of this study is to evaluate the extent and incidence of opaque bubble layer (OBL) using laser-assisted in situ keratomileusis (LASIK) flaps created with the Alcon/WaveLight® FS200 femtosecond laser as a result of a recent change in flap programming parameters aiming to reduce further the incidence and extent of OBL.
Methods:
Intraoperative digital images of flaps from 36 consecutive patients (72 eyes) subjected to bilateral femtosecond-assisted LASIK were analyzed using a proprietary computerized technique. The incidence and extent of OBL was measured and reported as a percentage of the entire flap area. Flap creation was performed with a 1.7 mm wide canal, implemented as an updated design intended to reduce the extent of OBL (group A). The same OBL parameters were investigated and compared in an age-matched and procedure-matched patients in whom the previous standard setting of a 1.3 mm wide canal was implemented (group B).
Results:
In group A, the average extent of OBL was 3.69% of the flap area (range 0%-11.34%). In group B, the respective values were 6.06% (range 0%-20.24%). We found the difference to be statistically significant (one-tailed P = 0.00452).
Conclusion:
This study suggests that there is a significant reduction in the incidence and extent of OBL when novel LASIK flap ventilation canal parameters of width and spot line separation are used.