The effects of particle size on measurement of airway hyperresponsiveness to methacholine
Nizar Naji1, Elaine Keung, Suzanne Beaudin
1Cardiorespiratory Research Laboratory, Department of Medicine, McMaster University, Hamilton, Ontario, Canada.
Background:
The effect of particle size on methacholine provocation concentration causing a decrease in forced expiratory volume of 1 second (FEV1) of 20% (PC20) is debatable.
Objective:
To evaluate the functional effects of 3 different particle size nebulizers on methacholine PC20.
Methods:
Participants were randomly assigned to have 3 methacholine challenges on 3 separate days. Nebulizer mass median aerodynamic diameter (MMAD) was provided by manufacturers. The Wright nebulizer (MMAD, 1.0 μm), Aeroneb (MMAD, 3 μm), and Aeroneb (MMAD, 5 μm) were calibrated, and the nebulizer outputs were calculated to administer 0.26 mL of methacholine over 120, 112, and 83 seconds, respectively. After each inhalation, spirometry was performed and the test was terminated when the PC20 was achieved.
Results:
Eight nonsmoking patients with mild asthma (4 male and 4 female) completed the study. The mean (SD) age was 25 (13.9) years, and the mean (SD) baseline FEV1 was 88% (11.3%). Patients using the Aeroneb (MMAD, 5 μm) nebulizer had the lowest PC20 (bronchoconstricted at lowest methacholine concentration), with a PC20 geometric mean of 0.62 mg/mL compared with patients using the Aeroneb (MMAD, 3.0 μm), who had a PC20 of 1.76 mg/mL, and patients using the Wright nebulizer (MMAD, 1.0 μm), who had a PC20 of 6.32 mg/mL. There was a significant difference in PC20 across all particle sizes (P < .001). The pairwise differences revealed a P < .001 between 3 μm and 1 μm and between 5 μm and 1 μm and a P = .008 between 5 μm and 3 μm.
Conclusion:
Our results reveal a variability in methacholine PC20 using 3 different nebulizers, despite adjusting the nebulizers' outputs. Our results are consistent with the previous reports, which recommended using larger particle size nebulizers in the assessment of airway hyperresponsiveness in asthma.
Trial Registration:
clinicaltrials.gov Identifier: NCT00529477.
Related Concept Videos
Chronic Obstructive Pulmonary Disease III: Chronic Bronchitis Features
Antiasthma Drugs: Methylxanthines
Theophylline is thought to inhibit phosphodiesterase enzymes, increasing intracellular levels of cyclic adenosine monophosphate (cAMP) and cyclic guanosine monophosphate (cGMP). This rise in cAMP and cGMP concentrations stimulates cardiac function,...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...