Related Experiment Video
Updated: May 11, 2026

Designing CAD/CAM Surgical Guides for Maxillary Reconstruction Using an In-house Approach
Published on: August 24, 2018
Preoperative analysis of the stability of fit of a patient-specific surgical guide
Joyce Van den Broeck1, Roel Wirix-Speetjens, Jos Vander Sloten
1a Department of Mechanical Engineering , KU Leuven, Celestijnenlaan 300C, Heverlee 3001 , Belgium.
Abstract:
Although the use of patient-specific surgical guides has gained popularity over the past decade, little research has been done to examine in an objective and qualitative way the fit of such instruments. In this study, we have developed a model to predict the stability of a guide designed to fit on a supporting bone surface, thereby providing feedback on the translational and rotational stability of the device. The method was validated by comparing different guide designs with respect to their stability on the contact surface and comparing these results to those measured with a set of experiments. This validation experiment indicates that our stability model can be used to predict the stability of the fit of a surgical guide during the preoperative design process.