An innovative distillation device for tritiated water analysis with high decontamination factor
Hsin-Fa Fang1, Chu-Fang Wang, Jeng-Jong Wang
1Institute of Nuclear Energy Research, Atomic Energy Council, No. 1000, Wunhua Road, Jiaan Village, Longtan Township, Taoyuan County 32546, Taiwan, ROC.
Abstract:
Institute of Nuclear Energy Research (INER) has designed an air-cooling distillation device and got a US patent. The decontamination factor (60)Co and (137)Cs is above 23,000. Tritium loss rate is one of testing items in ASTM D4107 Standard Test Method for Tritium in Drinking Water. In this study, the 3 levels (high, middle and low level) of tritium concentration of testing samples for the loss rate test were prepared similar to the concentrations reported in ASTM D4107. The loss rate of the high level is -2.37%, the middle is -2.31% and the low level is -2.47%. These results show that the air-cooling distillation device has good performance in the environmental water tritium analysis work.