Whole-body PET/MRI of Pediatric Patients: The Details That Matter
Published on: December 19, 2017
Comparison of standardized uptake values in normal structures between PET/CT and PET/MRI in an oncology patient
Sharif Kershah1, Sasan Partovi, Bryan J Traughber
1Department of Radiology, University Hospitals Case Medical Center, Case Western Reserve University, 11100 Euclid Avenue, Cleveland, OH, 44106, USA.
Purpose:
The purpose of this study was to compare and correlate standardized uptake values (SUV) derived from magnetic resonance attenuation correction (MRAC) with those derived from computed tomography attenuation correction (CTAC) in an oncology patient population.
Procedures:
The HIPAA-compliant study was approved by the Internal Review Board and all subjects gave written informed consent prior to inclusion in the study. Forty patients (mean age 61 ± 15.1; 20 male) referred for clinically indicated 2-deoxy-2-[(18)F]fluoro-D-glucose (FDG) positron emission tomography/computed tomography (PET/CT) scans also underwent a PET/magnetic resonance imaging (MRI) examination. MRAC was performed using an automatic three-segment model. Regions of interest were drawn over eight normal structures in order to obtain SUVmax and SUVmean values. Spearman rank correlation coefficients (r) were calculated and two-tailed paired t tests were performed to compare the SUVmax and SUVmean values obtained from CTAC with those from MRAC.
Results:
The mean time after FDG injection was 66 ± 7 min for PET/CT and 117 ± 15 min for PET/MRI examination. MRAC SUV values were significantly lower than the CTAC SUV values in mediastinal blood pool (p < 0.001 for both SUVmax and SUVmean) and liver (p = 0.01 for SUVmean). The MRAC SUV values were significantly higher in bone marrow (p < 0.001 for both SUVmax and SUVmean), psoas major muscle (p < 0.001 for SUVmax), and left ventricular myocardium (p < 0.001 for SUVmax and p = 0.01 for SUVmean). For the other normal structures, no significant difference was observed. When comparing SUV values generated from CTAC versus MRAC, high correlations between CTAC and MRAC were observed in myocardium (r = 0.96/0.97 for SUVmax/mean), liver (r = 0.68 for SUVmax), bone marrow (r = 0.80/0.83 for SUVmax/mean), lung tissue (r = 0.70 for SUVmax), and mediastinal blood pool (r = 0.0.68/.069 for SUVmax/mean). Moderate correlations were found in lung tissue (r = 0.67 for SUV mean), liver (r = 0.66 for SUVmean), fat (r = 0.48/0.53 for SUVmax/mean), psoas major muscle (r = 0.54/0.58 for SUVmax/mean), and iliacus muscle (r = 0.41 for SUVmax). Low correlation was found in iliacus muscle (r = 0.32 for SUVmean).
Conclusions:
Using the automatic three-segment model, our study showed high correlation for measurement of SUV values obtained from MRAC compared to those from CTAC, as the reference standard. Differences observed between MRAC and CTAC derived SUV values may be attributed to the time-delay between the PET/CT and PET/MRI scans or biologic clearance of radiotracer. Further studies are required to assess SUV measurements when performing different MR attenuation correction techniques.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...