Related Experiment Video
Updated: May 11, 2026

Development and Validation of an Ultrasensitive Single Molecule Array Digital Enzyme-linked Immunosorbent Assay for Human Interferon-α
Published on: June 14, 2018
Highly specific and sensitive measurements of human and monkey interleukin 21 using sequential protein and tryptic
Joe Palandra1, Alyce Finelli, Ming Zhu
1Pfizer Worldwide Research & Development, Andover, MA 01810, USA. joe.palandra@pfizer.com
Abstract:
A highly specific and sensitive immunoaffinity LC-MS/MS assay for quantification of human and cynomolgus monkey interleukin 21 (IL-21) was developed, qualified, and implemented. The workflow includes offline enrichment of IL-21 using an anti-IL-21 capture antibody, followed by isolation using magnetic beads, trypsin digestion, online enrichment of IL-21 derived tryptic peptides using antipeptide antibodies, and quantification using nanoflow LC-MS/MS. This assay was developed and qualified in human and cynomolgus monkey serum and tissues with a lower limit of quantitation of 0.78 pg/mL based on the intact cytokine. Both intra- and interbatch precision and accuracy, as well as stability and recovery, were found to be acceptable. IL-21 was not detected in serum from normal healthy volunteers or from autoimmune disease patients. However, IL-21 levels were quantified in cynomolgus monkey spleen and colon tissue and normal and inflammatory bowel disease (IBD) human colon tissue as well as hyperplasia human tonsils.