Yuejin Tong1, Haidong Li, Huaimin Guan
1College of Chemistry and Chemical Engineering, Fujian Normal University, Fuzhou 350007, China. tongyuejin@yahoo.com.cn
Abstract:
A monolithic molecular imprinting sensor based on ceramic carbon electrode (CCE) has been reported. The sensor can be renewed simply by smoothing. It was fabricated by thoroughly mixing multiwalled carbon nanotube@molecularly imprinted polymer (MWCNT@MIP), graphite powder, and silicon alkoxide, and then packing the resulting complex mixture of components firmly into the electrode cavity of a Teflon sleeve. The incorporated MWCNT@MIP in CCEs functioned as a recognition element for cholesterol determination. The MWCNT@MIP-CCEs were tested in the presence or absence of cholesterol by cyclic voltammetry and linear sweep voltammetry. The cholesterol sensor has excellent sensitivity with a linear range of 10-300nM and a detection limit of 1nM (S/N=3). The monolithic molecular imprinting sensor exhibits good stability, high sensitivity, and user-friendly reusability for cholesterol determination. This study shows that CCE is a promising matrix for MIP sensors.