Related Experiment Video
Updated: May 11, 2026

Assessing Binocular Central Visual Field and Binocular Eye Movements in a Dichoptic Viewing Condition
Published on: July 21, 2020
A device for continuous monitoring of true central fixation based on foveal birefringence
Boris Gramatikov1, Kristina Irsch, Marie Müllenbroich
1Division of Pediatric Ophthalmology and Adult Strabismus, The Wilmer Eye Institute, Johns Hopkins University School of Medicine, Wilmer 233, 600 N. Wolfe St, Baltimore, MD 21287, USA. bgramat@jhmi.edu
Abstract:
A device for continuous monitoring of central fixation utilizes birefringence, the property of the Henle fibers surrounding the human fovea, to change the polarization state of light. A circular scan of retinal birefringence, where the scanning circle encompasses the fovea, allows identification of true central fixation-an assessment much needed in various applications in ophthalmology, psychology, and psychiatry. The device allows continuous monitoring for central fixation over an extended period of time in the presence of fixation targets and distracting stimuli, which may be helpful in detecting attention deficit hyperactivity disorder, autism spectrum disorders, and other disorders characterized by changes in the subject's ability to maintain fixation. A proof-of-concept has been obtained in a small study of ADHD patients and normal control subjects.