The importance of commutability of reference materials used as calibrators: the example of ceruloplasmin
Ingrid Zegers1, Robert Beetham, Thomas Keller
1Institute for Reference Materials and Measurements (IRMM), Joint Research Centre, European Commission, Geel, Belgium. ingrid.zegers@ec.europa.eu
Background:
Different methods for ceruloplasmin tend to give different results in external quality assessment schemes. During the production of the certified reference material ERM-DA470k/IFCC discrepant measurement results were also found for ceruloplasmin measured with different methods, and consequently the protein could not be certified in the material.
Methods:
We performed a commutability study with 30 serum samples and the reference materials ERM-DA470, ERM-DA470k/IFCC, and ERM-DA472/IFCC, using 6 different methods. Data were analyzed according to the CLSI Guideline C53-A to assess whether the reference materials had the same behavior as the serum samples with respect to measurement results obtained with combinations of the methods used.
Results:
Measurement results from different methods showed a good linear correlation for the serum samples. ERM-DA470 showed marked noncommutability for certain combinations of methods. ERM-DA470k/IFCC and ERM-DA472/IFCC were commutable for more combinations of methods. The lack of commutability of ERM-DA470 for certain combinations of methods correlates with results from the UK National External Quality Assessment Service showing discrepancies between results from these methods. For serum stored in the presence of sodium azide the results from different methods are essentially equivalent.
Conclusions:
Ceruloplasmin in ERM-DA470 is a fully documented example of a situation in which, due to lack of commutability, the use of a common material for calibration did not lead to harmonization .
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...