Related Experiment Video
Updated: May 11, 2026

The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Is Montgomery tracheal Safe-T-Tube clinical failure induced by biofilm?
Massimo Fusconi1, Vincenza Rita Lo Vasco, Andrea Delfini
1Department of Sensory Organs, Sapienza University of Rome, Italy.
Objectives:
Montgomery Safe-T-Tube deterioration and early biofilm colonization may explain the discomfort claimed by many patients and clinical failures. The aim of the study was to analyze the deterioration of Montgomery Safe-T-Tube morphological and mechanical properties in vivo in 16 patients by using microbiological methods, optical and electron microscopy, and engineering tests.
Study Design:
Prospective controlled study at a single medical center.
Setting:
University hospital.
Subjects And Methods:
The study, conducted from April 2007 to February 2012 at the "Sapienza" University of Rome, was designed to collect 2 Montgomery Safe-T-Tubes from each patient. The first was removed 3 to 15 days after insertion (group A) and the second at least 90 days after (group B). Specimens underwent microbiologic assays, electron microscopic analysis, immunocytologic analysis, and mechanical tests.
Results:
Microorganisms were not isolated in 2 group A cases (12%), whereas they were in all group B cases. Biofilm was identified in 11 of 16 (69%) group A samples and in 16 of 16 (100%) group B samples (P = .0149) using scanning electron microscopy. Immunohistochemistry showed monocyte-granulocyte line cells producing interleukin-1β on the external surfaces of Montgomery Safe-T-Tubes. The tensile test showed that the wear related to the longer period of use makes Montgomery Safe-T-Tubes more rigid than newer ones.
Conclusion:
Early biofilm colonization takes place in Montgomery Safe-T-Tubes in most cases. The mechanical decay could be justified in part by the destructive biofilm activity and by the release of inflammatory effectors and enzymes.
More Related Videos
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of the...
Microbiota of the Respiratory Tract