High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Surface-activated chemical ionization-electrospray ionization mass spectrometry combined with two-dimensional serial
Matteo Conti1, Roberto Motta, Cristina Puggioli
1Central Laboratory, Department of Haematology, Oncology, and Laboratory Medicine, Sant'Orsola-Malpighi Hospital, Bologna, Italy. matteo.conti@teletu.it
Rationale:
Drug stability is an important quality-control issue for pharmaceutical and clinical practices. Among the analytical techniques that address this issue, liquid chromatography/mass spectrometry (LC/MS) can be very useful, especially when assessing the quality of liquid formulations, because it is intrinsically sensitive, selective, and a rapid analytical technique. However, LC/MS suffers from technical drawbacks, e.g., matrix effects, and the production of in-source degradation products, which can limit its usefulness.
Methods:
To overcome the aforementioned drawbacks associated with LC/MS, we introduce an innovative approach (2D-LC/SACI-ESI-MS/MS) that incorporates two-dimensional liquid chromatography (2D-LC) in conjunction with an MS system consisting of a surface-activated chemical ionization-electrospray ionization chamber (SACI-ESI), an ion trap MS analyzer, and tandem mass spectrometry.
Results:
To validate our 2D-LC/SACI-ESI-MS/MS system stability studies were performed on the computerized tomography contrast agents, iohexol, iodixanol, iopamidol, iomeprol, iopromide, and iobitridol, either alone or in binary combination. The matrix effects, in-source analyte degradation, and analytical performance were compared with those obtained using a one-dimensional LC/MS configuration. The accuracy coefficient of variance (CV) = 1-4%, and degradation (loss of water and other chemical moieties) was greatly reduced, attesting to the usefulness of this system for drug stability measurements.
Conclusions:
Our new approach improves the performance (sensitivity, accuracy, and signal stability) of LC/MS instrumentation for drug stability measurements by reducing signal suppression effects and in-source chemical reactions.
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Tandem Mass Spectrometry
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Ion-Exchange Chromatography
MALDI-TOF Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...