Published on: February 27, 2026
Thrombin-accelerated quick clotting serum tubes: an evaluation with 22 common biochemical analytes
1Department of Pathology, Clinical Biochemistry Laboratories, Singapore General Hospital, Outram Road, Singapore 169608.
Advances in Hematology
|May 9, 2013
Summary
BD Vacutainer Rapid Serum Tubes (RSTs) offer faster clotting times and comparable test results to Serum Separator Tubes (SSTs). This can improve laboratory efficiency without compromising accuracy for key biochemical markers.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Preanalytical Variables
Background:
- Clinical laboratories face increasing workloads and demand for faster test result reporting.
- Optimizing preanalytical variables is crucial for improving laboratory efficiency and performance.
- Current widely used Serum Separator Tubes (SSTs) have established performance benchmarks.
Purpose of the Study:
- To evaluate the performance of BD Vacutainer Rapid Serum Tubes (RSTs) in comparison to BD Vacutainer Serum Separator Tubes (SSTs).
- To assess test result comparability, clotting time, and serum analyte stability between RSTs and SSTs.
- To determine the potential of RSTs in reducing turnaround times in clinical laboratories.
Main Methods:
- A field study was conducted comparing BD Vacutainer RSTs and SSTs.
- Test performance was evaluated for renal, liver, lipid, cardiac, thyroid, and prostate biochemical markers.
- Clotting times and analyte stability at 4°C over 4 days were assessed in a study population of 160 patients and healthy subjects.
Main Results:
- Biochemical marker results from RSTs were comparable to those from SSTs.
- RSTs demonstrated significantly faster clotting times, with a median of 2.05 minutes.
- Analyte stability was maintained for 4 days at 4°C, with minor exceptions for bicarbonate, electrolytes, and albumin.
Conclusions:
- BD Vacutainer RSTs significantly reduce serum specimen clotting time compared to SSTs.
- The use of RSTs can lead to further reductions in overall blood collection to result reporting time.
- RSTs offer a viable alternative to SSTs, providing time savings without substantial compromise on test accuracy and performance.