A multicenter, prospective, randomized comparison of a novel signal transmission capsule endoscope to an existing
Eric H Choi1, Klaus Mergener, Carol Semrad
1Division of Gastroenterology and Hepatology, Riverside Medical Clinic, Riverside, California, USA.
Background:
MiroCam, a capsule endoscope, uses a novel transmission technology, electric-field propagation, which uses the human body as a conduction medium for data transmission.
Objective:
To compare the ability of the MiroCam (MC) and PillCam (PC) to identify sources of obscure GI bleeding (OGIB).
Design:
Prospective, multicenter, comparative study.
Setting:
Six academic hospitals.
Patients:
A total of 105 patients with OGIB.
Intervention:
Patients ingested both the MC and PC capsules sequentially in a randomized fashion.
Main Outcome Measurements:
Concordance of rates in identifying a source of OGIB, operational times, and rates of complete small-bowel examination.
Results:
Data analysis resulted in 43 (48%) "abnormal" cases identifying a source of OGIB by either capsule. Twenty-four cases (55.8%) were positive by both capsules. There was negative agreement in 46 of 58 cases (79.3%). The κ index was 0.547 (χ(2) = 1.32; P = .36). In 12 cases, MC positively identified a source that was not seen on PC, whereas in 7 cases, PC positively identified a source that was not seen on MC. MC had a 5.6% higher rate of detecting small-bowel lesions (P = .54). MC captured images at 3 frames per second for 11.1 hours, and PC captured images at 2 frames per second for 7.8 hours (P < .0001). Complete small-bowel examination was achieved in 93.3% for MC and 84.3% for PC (P = .10).
Limitations:
Readers were not blinded to the particular capsule they were reading.
Conclusion:
A positive diagnostic finding for OGIB was identified by either capsule in 48% of cases. The concordance rate between the 2 capsules was comparable to that of prior studies in identifying sources of small-bowel bleeding. The longer operational time of the MC may result in higher rates of complete small-bowel examination, which may, in turn, translate into a higher rate of detecting small-bowel lesions. (
Clinical Trial Registration Number:
NCT00878982.).
Related Concept Videos
Endoscopic Procedures III: Video Capsule Endoscopy
Urologic Endoscopic Procedure: Cystoscopic Examination
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due to...
Endoscopic Procedures V: ERCP
Patient...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to: