Guide wire shadow assessed by shading index is reduced in sparse spring coil wire in optical coherence tomography
Fumiaki Nakao1, Tooru Ueda, Shigehiko Nishimura
1Department of Cardiology, Yamaguchi Grand Medical Center, 77 Ohsaki, Hofu, Yamaguchi, 747-8511, Japan, nakao-ymghp@umin.ac.jp.
Abstract:
One of the major problems of a Fourier-domain optical coherence tomography (FD-OCT) is the guide wire (GW) shadow which disturbs precise coronary assessment. If two or more GWs are used in a bifurcation PCI, the GW shadow becomes larger. In FD-OCT, GWs were usually observed as a crescent shape, but GWs with the sparse spring coil were observed as either round or crescent shape. The measured angle making GW shadow of GW without the sparse spring coil was similar to its theoretical angle (30.4 ± 1.7° vs. 30.1 ± 0.7°, p = 0.21); however, the measured angle of GW with the sparse spring coil was significantly smaller than its theoretical angle (16.8 ± 4.8° vs. 28.7 ± 1.5°, p < 0.01). For standardization of shading-effect of GW, a virtual diameter of GW which was calculated from a measured angle, an actual diameter of GW, and a distance between an imaging catheter and GW was defined as the shading index. The shading index of GW with the sparse spring coil was significantly smaller than that of GW without the sparse spring coil (0.008 ± 0.002 vs. 0.014 ± 0.001, p < 0.01). Shading indices of GWs with the sparse spring coil were smaller than their actual diameters. The sparse spring coil structure could contribute to the reduction of the GW shadow in FD-OCT. Also, we proposed the new and useful shading index to indicate the shading-effect of GW.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography