Dynamic Navigation for Dental Implant Placement
Published on: September 13, 2022
Optimal caliper placement: manual vs automated methods
1Department of Obstetrics and Gynaecology, University of Tuebingen, Tuebingen, Germany.
Objective:
To examine the inter- and intra-operator repeatability of manual placement of callipers in the assessment of basic biometric measurements and to compare the results to an automated calliper placement system.
Methods:
Stored ultrasound images of 95 normal fetuses between 19 and 25 weeks' gestation were used. Five operators (two experts, one resident and two students) were asked to measure the BPD, OFD and FL two times manually and automatically. For each operator, intra-operator repeatability of the manual and automated measurements was assessed by within operator standard deviation. For the assessment of the interoperator repeatability, the mean of the four manual measurements by the two experts was used as the gold standard.The relative bias of the manual measurement of the three non-expert operators and the operator-independent automated measurement were compared with the gold standard measurement by means and 95% confidence interval.
Results:
In 88.4% of the 95 cases, the automated measurement algorithm was able to obtain appropriate measurements of the BPD, OFD, AC and FL. Within operator standard deviations of the manual measurements ranged between 0.15 and 1.56, irrespective of the experience of the operator.Using the automated biometric measurement system, there was no difference between the measurements of each operator. As far as the inter-operator repeatability is concerned, the difference between the manual measurements of the two students, the resident, and the gold standard was between -0.10 and 2.53 mm. The automated measurements tended to be closer to the gold standard but did not reach statistical significance.
Conclusion:
In about 90% of the cases, it was possible to obtain basic biometric measurements with an automated system. The use of automated measurements resulted in a significant improvement of the intra-operator but not of the inter-operator repeatability, but measurements were not significantly closer to the gold standard of expert examiners.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...