Related Experiment Video
Updated: May 11, 2026

Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing
Published on: June 1, 2012
Micro-drilling of polymer tubular ultramicroelectrode arrays for electrochemical sensors
Jan Kafka1, Steen Skaarup, Oliver Geschke
1Department of Micro- and Nanotechnology, DTU Nanotech, Technical University of Denmark, Ørsteds Plads, Building 345Ø, DK-2800 Kongens Lyngby, Denmark. jk@inmoldbiosystems.com
Abstract:
We present a reproducible fast prototyping procedure based on micro-drilling to produce homogeneous tubular ultramicroelectrode arrays made from poly(3,4-ethylenedioxythiophene) (PEDOT), a conductive polymer. Arrays of Ø 100 µm tubular electrodes each having a height of 0.37 ± 0.06 µm were reproducibly fabricated. The electrode dimensions were analyzed by SEM after deposition of silver dendrites to visualize the electroactive electrode area. The electrochemical applicability of the electrodes was demonstrated by voltammetric and amperometric detection of ferri-/ferrocyanide. Recorded signals were in agreement with results from finite element modelling of the system. The tubular PEDOT ultramicroelectrode arrays were modified by prussian blue to enable the detection of hydrogen peroxide. A linear sensor response was demonstrated for hydrogen peroxide concentrations from 0.1 mM to 1 mM.