In vivo evaluation of carotid artery stents using multi-detector-row computed tomography angiography: comparison of 3
Youn Joo Lee1, Yeon Soo Lim, Na Young Jung
1Department of Radiology, Bucheon St Mary's Hospital, College of Medicine, The Catholic University of Korea, Republic of Korea.
Objective:
The objective of this study was to assess the lumen visibility of carotid artery stents using multi-detector-row computed tomography (MDCT) angiography; compare medium-smooth, medium-sharp, and sharp kernels; and correlate these results to those of digital subtraction angiography (DSA).
Methods:
We retrospectively evaluated 52 stents from 51 patients who underwent 16- and 64-slice MDCT angiography. Lumen diameters were measured 3 times by 2 neuroradiologists, and artificial luminal narrowing was calculated. To assess detection of in-stent restenosis (>50%), 2 neuroradiologists evaluated all MDCT and DSA images.
Results:
Excellent intraobserver and interobserver agreements were obtained for the lumen diameter measurements (P < 0.001). Lumen diameter visibility improved, and artificial luminal narrowing decreased from the medium-smooth kernel to the sharp kernel. Visual estimation of all CT angiography using the 3 kernels showed high accuracy for detection of in-stent restenosis (>50%), compared with DSA.
Conclusions:
Computed tomography angiography using a sharp kernel allows for more accurate assessment of lumen visibility after carotid artery stenting.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System IV: CMRI