Simultaneous Determination of Miconazole Nitrate and Metronidazole in Different Pharmaceutical Dosage Forms by Gas
1Bioanalytical Laboratory, Department of Chemistry, Faculty of Science, University of Aleppo, Aleppo, Syria.
Abstract:
A simple, rapid and precise gas chromatographic method has been developed for the simultaneous determination of miconazole nitrate (MIZ) and metronidazole (MNZ) in tablets and ovules, using a capillary column AE.SE-54 (15 m × 0.53 mm, i.d.) and nitrogen as a carrier gas at a flow rate of 9 mL min(-1). The oven temperature was programmed at 140°C for 3 min, with a rise of 40°C min(-1) up to 180°C (held for 2 min) and then increased to a final temperature of 250°C. The injector and detector port temperatures were maintained at 260°C. Detection was carried out using flame ionization detector. Results of assay and recovery studies were statistically evaluated for its accuracy and precision. The retention times were about 3.50 and 12.90 min for MNZ and MIZ, respectively. Linearity ranges were 50.0-6030.0 and 62.5-2000.0 μg mL(-1) for MNZ and MIZ, with limit of detection values of 2.5 and 3.1 μg mL(-1), respectively. Correlation coefficients (R(2)) of the regression equations were greater than 0.999 in all cases. No interference from any components of pharmaceutical dosage forms or degradation products was observed. According to the validation results, the proposed method was found to be specific, accurate, precise and could be applied to the simultaneous quantitative analysis of MIZ and MNZ in tablets and ovules.
More Related Videos
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Automated Microbial Diagnostics