Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Calibration Curves: Correlation Coefficient01:10

Calibration Curves: Correlation Coefficient

In a linear calibration curve, there is a value called the calibration coefficient, denoted by 'r,' which measures the strength and the direction of association between two variables. The correlation coefficient value ranges from −1 to +1. A value of +1 indicates a perfect positive linear correlation, −1 denotes a perfect negative correlation, and 0 implies no correlation between the two variables. A positive correlation value establishes that as one variable increases, the other increases, and...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Apoptosis promotes fertility in <i>C. elegans</i> by maintaining functional germline morphology.

bioRxiv : the preprint server for biology·2025
Same author

Characterization or Monolithic Scintillation Detectors Etched with Laser Induced Optical Barriers.

IEEE transactions on radiation and plasma medical sciences·2021
Same author

Impact of event positioning algorithm on performance of a whole-body PET scanner using one-to-one coupled detectors.

Physics in medicine and biology·2018
Same author

Abstracts of the 33rd International Austrian Winter Symposium : Zell am See, Austria. 24-27 January 2018.

EJNMMI research·2018
Same author

Recent developments in time-of-flight PET.

EJNMMI physics·2016
Same author

Impact of detector design on imaging performance of a long axial field-of-view, whole-body PET scanner.

Physics in medicine and biology·2015
Same journal

Impact of apertures on the out-of-field secondary neutron dose in collimated proton pencil-beam scanning.

Physics in medicine and biology·2026
Same journal

Quantifying cardiac deformable image registration accuracy and its dosimetric variability for 4D dose accumulation in stereotactic arrhythmia radioablation.

Physics in medicine and biology·2026
Same journal

Probabilistic modelling of bilateral lymphatic spread in oral cavity squamous cell carcinoma for personalised elective nodal treatment.

Physics in medicine and biology·2026
Same journal

A Monte Carlo simulation tool to analyze breast cancer trial outcomes: application to FAST-Forward trial.

Physics in medicine and biology·2026
Same journal

Effective contrast-enhanced preprocessing for intracranial artery segmentation in digital subtraction angiography.

Physics in medicine and biology·2026
Same journal

Improving Plan Quality in Adaptive Proton Therapy Using an Interactive Dose Modification Tool.

Physics in medicine and biology·2026
See all related articles

Related Experiment Video

Updated: May 11, 2026

Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
10:21

Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function

Published on: August 8, 2019

TOF PET offset calibration from clinical data.

M E Werner1, J S Karp

  • 1Department of Radiology, The University of Pennsylvania, Philadelphia, PA 19104, USA. matt.werner@uphs.upenn.edu

Physics in Medicine and Biology
|May 21, 2013
PubMed
Summary
This summary is machine-generated.

This study introduces automated timing calibration for time-of-flight positron emission tomography (TOF PET) using existing data. This method enhances TOF PET imaging without specialized equipment, improving image quality and accessibility.

More Related Videos

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities
07:13

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities

Published on: October 27, 2023

Novel Quantification Protocol for Cardiovascular Calcification Progression Using Longitudinal MicroPET/MicroCT Images
08:02

Novel Quantification Protocol for Cardiovascular Calcification Progression Using Longitudinal MicroPET/MicroCT Images

Published on: November 15, 2024

Related Experiment Videos

Last Updated: May 11, 2026

Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
10:21

Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function

Published on: August 8, 2019

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities
07:13

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities

Published on: October 27, 2023

Novel Quantification Protocol for Cardiovascular Calcification Progression Using Longitudinal MicroPET/MicroCT Images
08:02

Novel Quantification Protocol for Cardiovascular Calcification Progression Using Longitudinal MicroPET/MicroCT Images

Published on: November 15, 2024

Area of Science:

  • Medical Imaging
  • Nuclear Medicine
  • Positron Emission Tomography

Background:

  • Accurate timing calibration is crucial for Time-of-Flight Positron Emission Tomography (TOF PET) image reconstruction.
  • Existing calibration methods often require specialized data acquisition protocols, limiting their flexibility and automation.

Purpose of the Study:

  • To present a novel, automated timing calibration technique for TOF PET.
  • To eliminate the need for dedicated calibration scans, enabling calibration using any clinical dataset.

Main Methods:

  • Developed a calibration method that integrates directly into the reconstruction process.
  • The technique utilizes arbitrary TOF PET datasets for calibration, allowing for retroactive application and use with existing data.

Main Results:

  • The proposed method provides fully automated timing calibration for TOF PET.
  • It can be applied to any clinical dataset, even retrospectively, without specialized acquisition.
  • Performance comparison demonstrates competitive results against existing calibration techniques.

Conclusions:

  • This automated calibration technique significantly improves the practicality and accessibility of TOF PET imaging.
  • It allows for improved image quality and correction of suboptimal or missing calibrations in retrospective datasets.