Management of respiration-induced motion with 4-dimensional computed tomography (4DCT) for pancreas irradiation
An Tai1, Zhiwen Liang, Beth Erickson
1Department of Radiation Oncology, Medical College of Wisconsin, Milwaukee, Wisconsin, USA. atai@mcw.edu
Purpose:
The purposes of this study were to quantify respiration-induced organ motions for pancreatic cancer patients and to explore strategies to account for these motions.
Methods And Materials:
Both 3-dimensional computed tomography (3DCT) and 4-dimensional computed tomography (4DCT) scans were acquired sequentially for 15 pancreatic cancer patients, including 10 randomly selected patients and 5 patients selected from a subgroup of patients with large tumor respiratory motions. 3DCTs were fused with 2 sets of 4DCT data at the end of exhale phase (50%) and the end of inhale phase (0%). The target was delineated on the 50% and 0% phase CT sets, and the organs at risk were drawn on the 3DCT. These contours were populated to the CT sets at other respiratory phases based on deformable image registration. Internal target volumes (ITV) were generated by tracing the target contours of all phases (ITV10), 3 phases of 0%, 20% and 50% (ITV3), and 2 phases of 0% and 50% (ITV2). ITVs generated from phase images were compared using percentage of volume overlap, Dice coefficient, geometric centers, and average surface distance.
Results:
Volume variations of pancreas, kidneys, and liver as a function of respiratory phases were small (<5%) during respiration. For the 10 randomly selected patients, peak-to-peak amplitudes of liver, left kidney, right kidney, and the target along the superior-inferior (SI) direction were 7.9 ± 3.2 mm, 7.1 ± 3.1 mm, 5.7 ± 3.2 mm, and 5.9 ± 2.8 mm, respectively. The percentage of volume overlap and Dice coefficient were 92% ± 1% and 96% ± 1% between ITV10 and ITV2 and 96% ± 1% and 98% ± 1% between ITV10 and ITV3, respectively. The percentage of volume overlap between ITV10 and ITV3 was 93.6 ± 1.1 for patients with tumor motion >8 mm.
Conclusions:
Appropriate motion management strategies are proposed for radiation treatment planning of pancreatic tumors based on magnitudes of tumor respiratory motions.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography