Related Experiment Video
Updated: May 11, 2026

Soil Lysimeter Excavation for Coupled Hydrological, Geochemical, and Microbiological Investigations
Published on: September 11, 2016
[Measurement and analysis of lead in soil using X-ray fluorescence spectroscopy]
Rong Zhang1, Yu-Jun Zhang, Wei Zhang
1Key Laboratory of Environment Optics and Technology, Chinese Academy of Sciences, Hefei 230031, China. rzhang@aiofm.ac.cn
Abstract:
The present paper analyzed the characteristics of X-ray fluorescence spectroscopy (XRF) of metal element lead in soil using the NITON XLt793 portable X-ray fluorescence spectra of heavy metal analyzer under laboratory conditions. The characteristic spectral lines of L(alpha) (energy: 10. 55 keV) and L(beta) (energy: 12. 61 keV) with different matrix elements were selected respectively for lead in the experiment. By measuring the intensities of the characteristic spectral line with different Pb concentration, the results demonstrate that the relation between concentration [mass fraction 10 x 10(-6) - 1 800 x 10(-6)] of Pb element and the intensity of the characteristic spectrum is well linear. The calibration curve of Pb was plotted based on the different concentration measurement results, and the limit of detection of 7.89 x 10(-6) was obtained for Pb in soil.
Related Concept Videos
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...