Published on: September 17, 2013
Operator auditory perception and spectral quantification of umbilical artery Doppler ultrasound signals
Ann Thuring1, K Jonas Brännström, Maria Ewerlöf
1Department of Obstetrics and Gynecology, Clinical Sciences Lund, Lund University, Lund, Sweden. ann.thuring@med.lu.se
Objective:
An experienced sonographer can by listening to the Doppler audio signals perceive various timbres that distinguish different types of umbilical artery flow despite an unchanged pulsatility index (PI). Our aim was to develop an objective measure of the Doppler audio signals recorded from fetoplacental circulation in a sheep model.
Methods:
Various degrees of pathological flow velocity waveforms in the umbilical artery, similar to those in human complicated pregnancies, were induced by microsphere embolization of the placental bed (embolization model, 7 lamb fetuses, 370 Doppler recordings) or by fetal hemodilution (anemia model, 4 lamb fetuses, 184 recordings). A subjective 11-step operator auditory scale (OAS) was related to conventional Doppler parameters, PI and time average mean velocity (TAM), and to sound frequency analysis of Doppler signals (sound frequency with the maximum energy content [MAXpeak] and frequency band at maximum level minus 15 dB [MAXpeak-15 dB] over several heart cycles).
Results:
WE FOUND A NEGATIVE CORRELATION BETWEEN THE OAS AND PI: median Rho -0.73 (range -0.35- -0.94) and -0.68 (range -0.57- -0.78) in the two lamb models, respectively. There was a positive correlation between OAS and TAM in both models: median Rho 0.80 (range 0.58-0.95) and 0.90 (range 0.78-0.95), respectively. A strong correlation was found between TAM and the results of sound spectrum analysis; in the embolization model the median r was 0.91 (range 0.88-0.97) for MAXpeak and 0.91 (range 0.82-0.98) for MAXpeak-15 dB. In the anemia model, the corresponding values were 0.92 (range 0.78-0.96) and 0.96 (range 0.89-0.98), respectively.
Conclusion:
Audio-spectrum analysis reflects the subjective perception of Doppler sound signals in the umbilical artery and has a strong correlation to TAM-velocity. This information might be of importance for clinical management of complicated pregnancies as an addition to conventional Doppler parameters.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of the Abdomen I: Inspection and Auscultation
The abdominal examination is a cornerstone of clinical medicine, serving as a critical tool in diagnosing various gastrointestinal (GI) diseases. It involves a systematic approach that includes inspection and auscultation, each with distinct yet complementary roles in assessing the abdomen. This article will delve into these two primary methods healthcare professionals use to examine the abdomen.
Inspection of the Abdomen
The first step in any abdominal examination is inspection.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Ultrasonography
During an ultrasonography procedure, a handheld device called a...