Non-invasive assessment of cardiac output at rest and during exercise by finger plethysmography
Christopher H Critoph1, Vimal Patel, Bryan Mist
1The Heart Hospital, Institute for Cardiovascular Science, University College London, London, UK. chriscritoph@doctors.org.uk
Aims:
We sought to determine the accuracy of finger plethysmography using pulse waveform analysis with brachial calibration for measurement of cardiac output during submaximal exercise by comparing it against an acetylene (C2H2) uptake technique.
Methods:
The study included 24 healthy volunteers (12 males, age 35 ± 8 years). Testing was performed on an upright cycle ergometer using an incremental protocol. Cardiac output measurements were performed at rest and during sub-maximal exercise using a single breath C2H2 uptake technique and continuously using finger plethysmography with brachial calibration.
Results:
Valid results at rest and during sub-maximal exercise were achieved in 20 of 24 participants. Cardiac output at rest was 5.3 ± 1.1 and 5.2 ± 1.2 l min(-1) for finger plethysmography and C2H2, respectively, P = 0.712. Mean difference between techniques was -0.1 ± 0.5 l min(-1). Cardiac output during submaximal exercise was 10.2 ± 2.3 and 10.3 ± 2.1 l min(-1) for finger plethysmography and C2H2, respectively, P = 0.898. Mean difference between techniques was 0.1 ± 1.5 l min(-1). The overall correlation between finger plethysmography and C2H2 data obtained during rest and exercise was r(2) = 0.872, P<0.0001. Mean rise in cardiac output during exercise was 4.9 ± 1.5 (finger plethysmography) and 5.1 ± 1.5 l min(-1) (C2H2), P = 0.64.
Conclusion:
Finger plethysmography determined cardiac output values both at rest and during sub-maximal exercise are comparable with those obtained using a single breath C2H2 uptake technique.
Related Concept Videos
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Pulse
The pulse serves as a clinical indicator...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for: