Lowering radiation exposure in CT angiography using automated tube potential selection and optimized iodine delivery
Florian Schwarz1, Katharina Grandl, Andreas Arnoldi
1Institute for Clinical Radiology, Hospital of the Ludwig Maximilian University of Munich, Campus Grosshadern, Marchioninistraße 15, 81377 Munich, Germany. Florian.Schwarz@med.uni-muenchen.de
Objective:
The purpose of this study was to determine the effectiveness of a radiation dose reduction strategy for CT angiography by the combination of higher iodine delivery rate and automated tube potential selection with adjusted reference values for tube current-exposure time product, as well as to measure the impact of this approach on image quality.
Subjects And Methods:
One hundred consecutive patients underwent high-pitch CT angiography of the thorax and abdomen using either 90 mL of iomeprol 300 (n = 44, protocol A) or 90 mL of iomeprol 400 (n = 56, protocol B) at the same flow rate. Automated tube potential selection was used with reference tube current-time products of 330 mAs and 250 mAs for protocols A and B, respectively. Twenty vascular segments were analyzed for attenuation and image noise by two readers. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated for all segments. The dose-length product (DLP) was documented to calculate effective dose and was compared between protocols both globally and for body mass index (BMI) subgroups.
Results:
There were no differences in sex, height, weight, or BMI between both groups. Images from patients scanned with protocol B showed higher levels of image noise but also higher signal in all vascular segments. As a result, there were no differences in SNR between both groups. Conversely, CNR was significantly higher for almost all vascular segments in the group scanned using protocol B. Furthermore, DLP was significantly lower when protocol B was used, particularly in patients with a BMI of less than 30.
Conclusion:
In CT angiography, a combination of higher iodine delivery rate and automated tube potential selection with adjusted reference values for the tube current-time product allows reductions in radiation dose by approximately 30% without compromising image quality.
More Related Videos
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...