Related Experiment Video
Updated: May 11, 2026

Sheathless Capillary Electrophoresis–Mass Spectrometry for Metabolic Profiling of Biological Samples
Published on: October 1, 2016
Column coupling isotachophoresis-capillary electrophoresis with mass spectrometric detection: characterization and
Pablo A Kler1, Tjorben N Posch, Martin Pattky
1Central Institute for Engineering, Electronics and Analytics, ZEA-3: Analytics, Forschungszentrum Jülich, Wilhelm-Johnen-Str., 52428 Jülich, Germany.
Abstract:
Two-dimensional electrophoretic separations are one of the most promising tools for the continuously growing needs of different bioanalytical fields such as proteomics and metabolomics. In this work we present the design and the implementation of a two-dimensional electrophoretic separation coupled to mass spectrometry. We started our work studying the sample transfer characteristics of different microfluidic interfaces compatible with capillary coupling for two-dimensional electrophoretic separations. These junctions are aimed at method decoupling and sample transfer in a modular two-dimensional electrophoretic separation system. In order to perform the characterization of the interfaces, we carried out capillary electrophoresis experiments and numerical simulations using three cationic compounds under different flow conditions. The comparison of the experimental and simulation results enables us to clearly define the desirable characteristics of interfaces in order to achieve method orthogonality with lossless sample transfer in a two-dimensional separation system. Finally, we present a glass microfluidic chip as interface for the implementation of a novel hybrid modular system for performing two-dimensional electrophoretic separations involving isotachophoresis and capillary electrophoresis. In this setup we include mass spectrometric and contactless capacitively coupled conductivity detection to monitor the separation process. We demonstrate the ability of the setup to be used as a flexible analysis tool by performing preconcentration, separation, detection and identification of four different human angiotensin peptides.
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Capillary Electrophoresis: Instrumentation
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Electrophoresis: Overview
There...
Ion-Exchange Chromatography
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: