Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Frameless high dose rate stereotactic lung radiotherapy: intrafraction tumor position and delivery time
Nicolas Peguret1, Max Dahele, Johan P Cuijpers
1Department of Radiation Oncology, VU University Medical Center, Amsterdam, The Netherlands.
Abstract:
Intrafraction change in tumor position (Δ) was evaluated for stereotactic lung radiotherapy delivered with flattening filter free volumetric modulated arc therapy. In 140 fractions from 32 patients mean Δ (±SD) was -0.7±1.4 mm (vertical), -0.7±1.3 mm (longitudinal) and +0.2±1.2 mm (lateral) with mean vector 2.1±1.2 mm. Mean delivery time was 4.4±3.4 min (mean beam-on 1.9±0.4 min).