Related Experiment Video
Updated: May 11, 2026

Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Preparation of 57Co point sources for the performance evaluation of nuclear imaging instruments
Sanjay Kumar Saxena1, Manoj Kumar, Yogendra Kumar
1Radiopharmaceuticals Division, Bhabha Atomic Research Centre, Trombay, Mumbai-400 085, Maharashtra, India.
Abstract:
This paper describes the utility of electrodeposition method to prepare (57)Co point sources. A description of the electrolytic cell, the process of (57)Co electrodeposition, encapsulation and quality control of the sealed (57)Co sources is presented. Sources containing ~3.7-4.81 MBq (0.10-0.13 mCi) of (57)Co were prepared, encapsulated in miniature titanium capsules and subjected to quality control tests to ensure compliance with regulatory requirements. The encapsulated (57)Co sources prepared by the reported procedure were found to be effective in routine performance evaluation of nuclear medicine instruments.