Intermittent subglottic secretion drainage may cause tracheal damage in patients with few oropharyngeal secretions
1Intensive Care Department, University Hospital, Vrije Universiteit Brussel, Laarbeeklaan, 101, B-1090 Brussels, Belgium(1).
Objective:
Injurious prolapse of tracheal mucosa into the suction port has been reported in up to 50% of intubated patients receiving continuous aspiration of subglottic secretions. We investigated whether similar injury could be inflicted by automated intermittent aspiration.
Methods:
Six consecutive patients, intubated with the Mallinckrodt TaperGuard Evac™ endotracheal tube, were studied. A flow sensor was placed between the vacuum regulating system and the mucus collector. Intermittent suctioning was performed at a pressure of -125 mmHg with a 25s interval and duration of 15s. After 24h, a CT scan of the tracheal region was performed.
Results:
Excessive negative suction pressure, a fast drop in aspiration flow to zero, and important "swinging" movements of secretions in the evacuation line were observed in all patients. Oral instillation of antiseptic mouthwash restored normal aspiration flow and secretion mobility. CT imaging showed marked entrapment of tracheal mucosa into the suction port in all patients.
Conclusion:
In patients with few oropharyngeal secretions, automated intermittent subglottic aspiration may result in significant and potential harmful invagination of tracheal mucosa into the suction lumen. A critical amount of fluid must be present in the oropharynx to assure adequate and safe aspiration.
Related Concept Videos
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of the...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...