Estimation of biological variation and reference change value of glycated hemoglobin (HbA(1c)) when two analytical
Fatma Ucar1, Gonul Erden, Zeynep Ginis
1Department of Clinical Biochemistry, Dışkapı Yıldırım Beyazıt Training and Research Hospital, Ankara, Turkey.
Objectives:
Available data on biological variation of HbA1c revealed marked heterogeneity. We therefore investigated and estimated the components of biological variation for HbA1c in a group of healthy individuals by applying a recommended and strictly designed study protocol using two different assay methods.
Design And Methods:
Each month, samples were derived on the same day, for three months. Four EDTA whole blood samples were collected from each individual (20 women, 9 men; 20-45 years of age) and stored at -80°C until analysis. HbA1c values were measured by both high performance liquid chromatography (HPLC) (Shimadzu, Prominence, Japan) and boronate affinity chromatography methods (Trinity Biotech, Premier Hb9210, Ireland). All samples were assayed in duplicate in a single batch for each assay method. Estimations were calculated according to the formulas described by Fraser and Harris.
Results:
The within subject (CV(I))-between subject (CV(G)) biological variations were 1.17% and 5.58%, respectively for HPLC. The calculated CV(I) and CV(G) were 2.15% and 4.03%, respectively for boronate affinity chromatography. Reference change value (RCV) for HPLC and boronate affinity chromatography was 5.4% and 10.4% respectively and individuality index of HbA(1c) was 0.35 and 0.93 respectively.
Conclusions:
This study for the first time described the components of biological variation for HbA1c in healthy individuals by two different assay methods. Obtained findings showed that the difference between CV(A) values of the methods might considerably affect RCV. These data regarding biological variation of HbA(1c) could be useful for a better evaluation of HbA(1c) test results in clinical interpretation.
Related Concept Videos
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
Bioequivalence Data: Statistical Interpretation
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...