Context:
Tracheal intubation in nonstandardised positions is associated with a higher risk of tube misplacement and may have deleterious consequences for patients. Video laryngoscopes for tracheal intubation facilitate both glottic view and success rates. However, their use in the ice-pick position has not been evaluated.
Objective:
To evaluate the role of video laryngoscopes for tracheal intubation in the ice-pick position.
Design:
A randomised, controlled manikin trial.
Setting:
A standardised airway manikin was placed in the corner of a room. Tracheal intubation was only possible from the lower right side of the manikin. In randomised order, participants used a standard Macintosh laryngoscope and GlideScope Ranger, Storz C-MAC, Pentax AWS, Airtraq and McGrath Series5 video laryngoscopes. Statistical analysis was performed using the Wilcoxon signed-rank and McNemar's tests; A P value of less than 0.05 was deemed statistically significant.
Participants:
Twenty anaesthesiologists, all emergency medicine board-certified.
Main Outcome Measures:
Time to first ventilation (primary); time to glottic view and confirmation of tube position (secondary).
Results:
Successful ventilation was achieved most rapidly with the Macintosh laryngoscope (36.1 ± 13.4 s; reference method), followed by the Airtraq (38.4 ± 36.3 s; P = n.s.), Pentax AWS (51.6 ± 43.3 s; P = n.s.) and Storz C-Mac (62.7 ± 49.7 s; P = n.s.). The use of the GlideScope Ranger (79.8 ± 61.9 s, P = 0.01) and McGrath series5 (79.8 ± 58.5 s, P = 0.023) resulted in significantly longer times. When comparing overall intubation success, the rate of successful tracheal intubation was higher with the Airtraq than with the McGrath Series5 (P = 0.031; all others n.s.).
Conclusion:
The use of video laryngoscopes did not result in higher success rates or faster tracheal intubation in the ice-pick position when compared with conventional laryngoscopy in this manikin study.
Trial Registration:
www.clinicaltrials.gov, NCT01210105.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Cardiopulmonary Resuscitation II: ACLS Airway Management