Electrical impedance tomography for verification of correct endotracheal tube placement in paediatric patients: a
D Steinmann1, M Engehausen, B Stiller
1Department of Occupational Medicine, University Medical Center Freiburg, Freiburg, Germany. daniel.steinmann@uniklinik-freiburg.de
Background:
Endotracheal tubes (ETTs) are frequently used in paediatric anaesthesia. Correct placement is crucial. The aim of this study was to evaluate electrical impedance tomography (EIT) for guiding and confirmation of paediatric ETT placement. In a retrospective analysis of stored EIT data, distribution of ventilation between left and right lung was used to verify correct paediatric ETT placement.
Methods:
Left and right lung ventilation was studied by EIT in 18 paediatric patients (median age: 53 months) requiring anaesthesia and endotracheal intubation. EIT was recorded before induction of anaesthesia, during mask ventilation, during ETT placement (including deliberate mainstem intubation), and after ETT repositioning according to the formula: ETT intubation depth (cm) = 3× ETT internal diameter (mm) or the mainstem intubation method (withdrawing the ETT 2 cm). Final ETT position was confirmed by fluoroscopy.
Results:
Following deliberate mainstem intubation, distribution of ventilation to the right lung was unequivocally demonstrated by EIT. Homogeneous distribution of ventilation between left and right lung monitored with EIT correlated in each patient with correct endotracheal ETT placement. The distribution of left and right lung ventilation differed significantly (P < 0.05) between the initial two-lung ventilation and subsequent right one-lung ventilation, and between right one-lung and subsequent two-lung ventilation according to auscultation and the final ETT position, respectively. In one patient, ETT was misplaced within the oesophagus which was also obvious from the EIT record.
Conclusion:
This study demonstrates that EIT enables non-invasive recognition of correct ETT placement. Homogeneous right-left-lung ventilation is an indicator for correct ETT placement.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...