Related Experiment Video
Updated: May 10, 2026

Rotating the Intraocular Lens to Prevent Posterior Capsular Opacification in Cataract Surgeries
Published on: July 7, 2023
[Computer-assisted phacoemulsification for hard cataracts]
M Zemba1, Adriana-Camelia Papadatu, Laura-Nicoleta Sîrbu
1Spitalul Universitar de Urgenţa Militar Central Dr. Carol Davila, Bucureşti.
Purpose:
to evaluate the efficiency of new torsional phacoemulsification software (Ozil IP system) in hard nucleus cataract extraction.
Methods:
45 eyes with hard senile cataract (degree III and IV) underwent phacoemulsification performed by the same surgeon, using the same technique (stop and chop). Infiniti (Alcon) platform was used, with Ozil IP software and Kelman phaco tip miniflared, 45 degrees. The nucleus was split into two and after that the first half was phacoemulsificated with IP-on (group 1) and the second half with IP-off (group 2). For every group we measured: cumulative dissipated energy (CDE), numbers of tip closure that needed manual desobstruction the amount of BSS used.
Results:
The mean CDE was the same in group 1 and in group 2 (between 6.2 and 14.9). The incidence of occlusion that needed manual desobstruction was lower in group 1 (5 times) than in group 2 (13 times). Group 2 used more BSS compared to group 1.
Conclusion:
The new torsional software (IP system) significantly decreased occlusion time and balanced salt solution use over standard torsional software, particularly with denser cataracts.
More Related Videos
Related Concept Videos
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...