Related Experiment Video
Updated: May 10, 2026

Application and Methodology of the Non-destructive 19F Time-domain NMR Technique to Measure the Content in Fluorine-containing Drug Products
Published on: August 22, 2017
A validated HPTLC method for estimation of moxifloxacin hydrochloride in tablets
Vandana Dhillon1, Alok Kumar Chaudhary
1Department of Pharmacy, IEC College of Engineering and Technology, Plot -4, Greater Noida, UP, India.
Abstract:
A simple HPTLC method having high accuracy, precision and reproducibility was developed for the routine estimation of moxifloxacin hydrochloride in the tablets available in market and was validated for various parameters according to ICH guidelines. moxifloxacin hydrochloride was estimated at 292 nm by densitometry using Silica gel 60 F254 as stationary phase and a premix of methylene chloride: methanol: strong ammonia solution and acetonitrile (10:10:5:10) as mobile phase. Method was found linear in a range of 9-54 nanograms with a correlation coefficient >0.99. The regression equation was: AUC = 65.57 × (Amount in nanograms) + 163 (r(2) = 0.9908).
More Related Videos
08:54Validated LC-MS/MS Panel for Quantifying 11 Drug-Resistant TB Medications in Small Hair Samples
Published on: May 19, 2020
08:57Sample Extraction and Simultaneous Chromatographic Quantitation of Doxorubicin and Mitomycin C Following Drug Combination Delivery in Nanoparticles to Tumor-bearing Mice
Published on: October 5, 2017