Development and validation of UV-Visible spectrophotometric baseline manipulation methodology for simultaneous
Vishnu P Choudhari1, Sanket R Parekar, Subhash G Chate
1Department of Pharmaceutical Analysis and Quality Assurance, Maharashtra Institute of Pharmacy, Kothrud, Pune, India.
Introduction:
A simple, economical, precise, and accurate new UV spectrophotometric baseline manipulation methodology for simultaneous determination of drotaverine (DRT) and etoricoxib (ETR) in a combined tablet dosage form has been developed.
Materials And Methods:
The method is based on baseline manipulation (difference) spectroscopy where the amplitudes at 274 and 351 nm were selected to determine ETR and DRT, respectively, in combined formulation and methanol was used as solvent. Both the drugs obey Beer's law in the concentration ranges of 4-20 μg/mL for DRT and 4.5-22.5 μg/mL for ETR.
Results:
The results of analysis have been validated statistically and recovery studies confirmed the accuracy and reproducibility of the proposed method which were carried out by following the ICH guidelines.
Conclusion:
It has been concluded that a new simple and accurate UV spectrophotometric baseline manipulation method was developed for simultaneous do not declare DRT and ETR in a combined tablet dosage form has been developed.
More Related Videos
08:57Sample Extraction and Simultaneous Chromatographic Quantitation of Doxorubicin and Mitomycin C Following Drug Combination Delivery in Nanoparticles to Tumor-bearing Mice
Published on: October 5, 2017
05:03Quantitative Analysis of Dietary Vitamin A Metabolites in Murine Ocular and Non-Ocular Tissues Using High-Performance Liquid Chromatography
Published on: December 27, 2024
Related Concept Videos
Therapeutic Drug Monitoring: Drug Analysis Methods
In Vitro Drug Release Testing: Overview, Development and Validation
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Drug Dissolution: Requirements and Profile Comparison
Drug Product Performance: In Vitro–In Vivo Correlation
Clinically Relevant Drug Product Specifications: Methods of Establishment